Gbs-elektronik MCA166-USB Manual de usuario Pagina 62

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 61
MCA166-USB
MCA Hardware Setup Commands:
Command name
CMD_SET_GAIN
Description The command sets the amplifier coarse(cg) and fine gain(fg).
cg: 2, 5, 10, 20, 50, 100, 200, 500 or 1000
fg: 5000 ... 15000 (for cg = 1000 fg: 5000 ... 10000)
Command syntax
Format integer integer integer long integer
Parameter preamble command cg fg end flag
Byte String (HEX) A5 5A 4C 00 cg
l
cg
h
fg
l
fg
h
00 00 B9 9B
Used setup value None
Affected setup value coarse_gain, fine_gain
MCA state STATE_READY
Command name
CMD_SET_INPUT_POLARITY
Description The command sets the input polarity for the internal amplifier.
ip = 0 positive input signals
ip = 1 negative input signals
Command syntax
Format integer integer integer long integer
Parameter preamble command ip 0 end flag
Byte String (HEX) A5 5A 56 00 ip 00 00 00 00 00 B9 9B
Used setup value None
Affected setup value mca_input_pol
MCA state STATE_READY
Command name
CMD_SET_MCA_INPUT
Description The command sets the ADC input to amplifier or to direct.
ip = 0 input amplifier
ip = 3 input direct (0 ... +3V, PUR OFF)
ip = 4 input direct (0 ... -3V, PUR OFF)
Command syntax
Format integer integer integer long integer
Parameter preamble command ip 0 end flag
Byte String (HEX) A5 5A 54 00 ip 00 00 00 00 00 B9 9B
Used setup value None
Affected setup value mca_input_adc
MCA state STATE_READY
62
Vista de pagina 61
1 2 ... 57 58 59 60 61 62 63 64 65 66 67 ... 71 72