Command name
CMD_SET_PUR
Description The command sets the pile up rejection.
Command syntax
pur = 0 turns PUR off
pur 0 turns PUR on
Format integer integer integer long integer
Parameter preamble command pur 0 end flag
Byte String (HEX) A5 5A 53 00 pur 00 00 00 00 00 B9 9B
Used setup value None
Affected setup value mca_pur
MCA state STATE_READY
Command name CMD_SET_FAST
Description This command set the fast discriminator threshold (0 ... 2499, default 400).
Command syntax
Format integer integer integer long integer
Parameter Preamble command fast 0 end flag
Byte String (HEX) A5 5A 50 00 fast 00 00 00 00 00 B9 9B
Used setup value None
Affected setup value Fast threshold
MCA state STATE_READY
Command name CMD_SET_SLOW
Description This command set the slow discriminator threshold (0 ... 2499, default 400).
Command syntax
Format integer integer integer long integer
Parameter Preamble command slow 0 end flag
Byte String (HEX) A5 5A 51 00 slow 00 00 00 00 00 B9 9B
Used setup value None
Affected setup value Slow threshold
MCA state STATE_READY