MCA166-USB
MCA Data Setup Commands:
Command name
CMD_SET_TDF
Description
The command sets the TDF parameter.
tdf: TDF(100 ... 3000, default 800)
Command syntax
Format integer integer integer long integer
Parameter preamble command tdf 0 end flag
Byte String (HEX) A5 5A 61 00 tdf
l
tdf
h
0 0 0 0 B9 9B
Used setup value None
Affected setup value tdf
MCA state STATE_READY
Command name
CMD_SET_UF6_ROI
Description
The command sets the begin and the end of a ROI used by the other UF6 commands.
r: ROI number
b: LLD<= b < e
e: b < e <= ULD, (e-b) < 500
Command syntax
Format integer integer integer integer integer integer
Parameter Preamble command r b e end flag
Byte String (HEX) A5 5A 64 00 r 00 b
l
b
h
e
l
e
h
B9 9B
Used setup value LLD, ULD
Affected setup value roi_begin[n], roi_end[n]
MCA state STATE_READY
Command name
CMD_SET_USER_DATA
Description
The command stores a 32 bit value(f) in the MCA parameter memory(64 entries).
e: 0 ... 63 parameter number
p: 0 ... long or float parameter
Command syntax
Format integer integer integer long or float integer
Parameter preamble command e P end flag
Byte String (HEX) A5 5A 57 00 e 00 p
l
p
l
p
h
p
h
B9 9B
Used setup value None
Affected setup value None
MCA state STATE_READY