Description of the MCA527 Firmware Commands
Command name CMD_SET_MEASURE_PZC
Compatibility Compatible to the MCA166 command.
Execution right Necessary
Description The command performs the Pole Zero Cancellation.
pv: 0 ... 2499 PZC control voltage
pm = 0 set PCZ only
pm > 0 set PZC and measure PZC offset
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer integer integer integer
Parameter preamble command pm pv 0 end flag
Byte string (HEX) A5 5A 58 00 pm
l
pm
h
pv
l
pv
h
00 00 B9 9B
Result data array
Byte offset 0 unused 106 bytes
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 unused 12 bytes
Byte offset 126 Checksum unsigned short
Byte offset 128 Number of measured pulses unsigned short
Byte offset 130 Averaged negative offset of measured input
pulses
short
Remarks This command takes at least 800 milliseconds to return, because it runs a measurement of this
duration.
Command name
CMD_SET_PZC_TIME_OFFSET
Compatibility
Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software
Execution right Necessary
Description The command sets the time offset for Pole Zero Cancellation.
t1: 0 ... 31 Time offset for low shaping time
t2: 0 ... 31 Time offset for high shaping time
Format integer integer integer integer integer integer
Parameter preamble command t1 t2 0 end flag
Byte string (HEX) A5 5A 60 00 t1 00 t2 00 00 00 B9 9B
15