MCA527
Command name CMD_SET_AHRC_PARAM
Compatibility New MCA527 command (since firmware version 14.00).
Execution right Necessary
Description The command allows to set the AHRC parameters.
param: parameter to be set
-1 = AHRC trigger threshold (allowed range = 0 … 65535
22
)
0 … 9 = AHRC [0 … 9] group width (allowed range = 1 …2147483647
22
)
value: new value corresponding to the allowed range
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer long integer
Parameter preamble command param value end flag
Byte string (HEX) A5 5A 2D 01 p
l
p
h
value
l
... ... value
h
B9 9B
22 The given intervals for the parameters only prevent a firmware hangup. They do not guarantee a meaningful result.