MCA527
The Ethernet communication uses the UDP protocol with the destination port 50000. The preamble is
prefixed 2 bytes (0xA5, 0x5A) for intern byte alignment within the MCA527.
The MCA527 can be configured to use a fixed customized IP address or to obtain the IP address
automatically from a DHCP server or if not available, to use Zero Configuration Networking (also named
Automatic Private IP Addressing).
While the MCA527 is writing a file, only the query commands and CMD_STOP command are executable. All
other commands are ignored and response with an error value. The file writing state is returned by
CMD_QUERY_STATE527_EX.
MCA Reset Command:
Command name
CMD_INIT
Compatibility Identical to the MCA166 command.
Execution right Necessary
Description All MCA Parameters will be reset to their initial state and the spectra are cleared. The measurement
is aborted. The preamplifier power and the high voltage are turned off!
Format integer integer integer long integer
Parameter preamble command 0 0 end flag
Byte string (HEX) A5 5A 41 00 00 00 00 00 00 00 B9 9B
MCA Clear Command:
Command name CMD_CLEAR
Compatibility Differs to the MCA166 command.
Execution right Necessary
Description Depending on the parameter: CLEAR_MEASUREMENT_DATA = 0 or 1
CLEAR_ROI = 2
CLEAR_ALL = 3
the command clears the measurement data, the ROI limits exclusively or all together. Before the
measurement data are cleared, a running measurement is immediately stopped.
Format integer integer integer long integer
Parameter preamble command clear 0 end flag
Byte string (HEX) A5 5A 44 00 clear 00 00 00 00 00 B9 9B