Gbs-elektronik MCA-527 Firmware Commands Manual de usuario Pagina 41

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 54
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 40
Description of the MCA527 Firmware Commands
Command name CMD_QUERY_STATE527_EX2
Compatibility New MCA527 command (since firmware version 14.00).
Execution right Not necessary
Description The command reads the state.
Format integer integer integer long integer
Parameter preamble command 0 0 end flag
Byte string (HEX) A5 5A 2F 01 00 00 00 00 00 00 B9 9B
Result data array
Byte offset 0 AHRC group 0 width unsigned long
Byte offset 4 AHRC group 1 width unsigned long
Byte offset 8 AHRC group 2 width unsigned long
Byte offset 12 AHRC group 3 width unsigned long
Byte offset 16 AHRC group 4 width unsigned long
Byte offset 20 AHRC group 5 width unsigned long
Byte offset 24 AHRC group 6 width unsigned long
Byte offset 28 AHRC group 7 width unsigned long
Byte offset 32 AHRC group 8 width unsigned long
Byte offset 36 AHRC group 9 width unsigned long
Byte offset 40 AHRC trigger threshold unsigned short
Byte offset 42 unused 2 bytes
Byte offset 44
Time window 0 width for gating mode 'sort by time'
50
unsigned long
Byte offset 48
Time window 1 width for gating mode 'sort by time'
50
unsigned long
Byte offset 52
Time window 2 width for gating mode 'sort by time'
50
unsigned long
Byte offset 56
Time window 3 width for gating mode 'sort by time'
50
unsigned long
Byte offset 60
Time window 4 width for gating mode 'sort by time'
50
unsigned long
Byte offset 64
Time window 5 width for gating mode 'sort by time'
50
unsigned long
Byte offset 68
Time window 6 width for gating mode 'sort by time'
50
unsigned long
Byte offset 72
Time window 7 width for gating mode 'sort by time'
50
unsigned long
Byte offset 76 unused 30 bytes
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 unused 12 bytes
Byte offset 126 Checksum unsigned short
Byte offset 128 MCA state unsigned short See CMD_QUERY_POWER
Byte offset 130 unused 2 bytes
50 Since firmware version 14.02.