Gbs-elektronik MCA-527 Firmware Commands Manual de usuario Pagina 54

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 54
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 53
MCA527
Name
CMD_QUERY_ADJUSTMENT_TABLE
Compatibility New MCA527 command.
Execution right Not necessary
Description The command returns the ADC values that has been measured with different settings at open signal
input. These data can be used to interpolate the ADC value which corresponds to the ground
line. A value is valid if it is unequal to -1.
Format integer integer integer long integer
Parameter Preamble command 0 0 end flag
Byte string (HEX) A5 5A 26 01 00 00 00 00 00 00 B9 9B
Result data array
Byte offset 0
measured with coarse gain = 2, offset DAC = 1024
short
Byte offset 2
measured with coarse gain = 2, offset DAC = 15360
short
Byte offset 4
measured with coarse gain = 5, offset DAC = 1024
short
Byte offset 6
measured with coarse gain = 5, offset DAC = 15360
short
Byte offset 8
measured with coarse gain = 10, offset DAC = 1024
short
Byte offset 10
measured with coarse gain = 10, offset DAC = 15360
short
Byte offset 12
measured with coarse gain = 20, offset DAC = 1024
short
Byte offset 14
measured with coarse gain = 20, offset DAC = 15360
short
Byte offset 16
measured with coarse gain = 50, offset DAC = 1024
short
Byte offset 18
measured with coarse gain = 50, offset DAC = 15360
short
Byte offset 20
measured with coarse gain = 100, offset DAC = 1024
short
Byte offset 22
measured with coarse gain = 100, offset DAC = 15360
short
Byte offset 24 unused 82 bytes
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 unused 12 bytes
Byte offset 126 Checksum unsigned short
Byte offset 128 MCA state unsigned short See CMD_QUERY_POWER
Byte offset 130 unused 2 bytes