MCA527
Command name CMD_SET_ROI
Compatibility Identical to the MCA166 command.
Execution right Necessary
Description The command sets the begin and end of the ROI for the preset integral and area.
(LLD <= begin < end and LLD < end <= ULD)
Format integer integer integer integer integer integer
Parameter preamble command beg end end flag
Byte string (HEX) A5 5A 49 00 beg
l
beg
h
end
l
end
h
00 00 B9 9B
Command name CMD_SET_REPEAT
Compatibility Identical to the MCA166 command.
Execution right Necessary
Description The command sets the number (rep: 0 ... 65535) of sweeps for repetitive measurement.
Rep = 0 means infinite repetitions.
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer long integer
Parameter preamble command rep 0 end flag
Byte string (HEX) A5 5A 4A 00 rep
l
rep
h
00 00 00 00 B9 9B
Command name
CMD_SET_MCS_CHANNEL
Compatibility Compatible to the MCA166 command.
Execution right Necessary
Description The command sets the number of channels (ch: 1 ... 16384) for MCS mode.
The command is ignored and responds with an error value if a measurement is still running.
Command syntax
Format integer integer integer long integer
Parameter preamble command ch 0 end flag
Byte string (HEX) A5 5A 63 00 ch
l
ch
h
00 00 00 00 B9 9B
Command name
CMD_SET_TIME_PER_CHANNEL
Compatibility Identical to the MCA166 command.
Execution right Necessary
Description The command sets the dwell time per channel (tpc: 1 ... 65535) for the MCS mode. The dwell time is
tpc * 10 ms.
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer long integer
Parameter preamble command tpc 0 end flag
Byte string (HEX) A5 5A 4B 00 tpc
l
tpc
h
00 00 00 00 B9 9B
Remarks See next command.